Measurement Specialties Inc. Model 9116 User’s Manual
Page 81 www.MEAS-SPEC.com
Transducer Coefficients Description
Re-zero Cal Adjustment (offset) term (Note 1)
Span Cal Adjustment (gain) term (Note 2)
User Defined Field (Note 4)
Transducer Manufacturing Reference Number (Note 5)
Transducer Full-Scale Range Code (See Appendix F) (Note 5)
(Note 1) Related command ‘w08’ can be used to download the offset term to the
sensor’s non-volatile memory (digitally or non-digitally compensated sensor).
(Note 2) Related command ‘w09’ can be used to download the gain term to the
sensor’s non-volatile memory (digitally or non-digitally compensated sensor).
(Note 4) Data is immediately stored to the sensor’s non-volatile memory.
The User Defined Date field (cc=07) is a 32-bit integer. Possible uses are to I
ndicate the date of last user zero and/or span calibration or possibly the date of
next required calibration. If this optional field is used, the user is responsible for
correctly encoding the date into the appropriate 32-bit integer value e.g., a
decimally encoded ‘yymmdd’ date. Any modifications of this field (using the
Download Internal Coefficients (‘v’) command) will result in the new value
automatically being entered to transducer non-volatile memory.
A special single Other Coefficients array is selected with array index aa=11
(hex). All the valid coefficient indexes (for this array only) are listed in the
following table:
Other Coefficients Description
EU Pressure Conversion scaler (default=1.0)