Measurement Specialties Inc. Model 9116 User’s Manual
Page 64 www.MEAS-SPEC.com
Example:
Send TCP/IP command to a Model 9116 module (via its open socket) to
calculate and set new offset coefficients for channels 16 through 13.
hF000
Read response, containing all new offset values (also stored in the module’s
volatile main memory):
0.0010 0.0020 0.0015 0.0025
Actual offset values are returned in the above response as decimal fixed-point
ASCII strings, each preceded by a space character. From left-to-right: they are
for channels 16, 15, 14, and 13.